Trinity Operating Procedures - TOPs
TOP-CALIB: Calibration Control Program
Rev: 2/23/2023
Scope: This procedure covers the calibration control program for precision measuring instruments. Instruments2 to be used for ANY measurement affecting the quality of Trinity Forge products must be entered into the Calibration System 4 and tagged with a current calibration sticker. This also includes gages calibrated by an outside ISO17025 lab.
Procedure:
At
the start
of the each shift: 1.
Verify1
that each instrument and/or
standard to be used is tagged with a current Trinity Forge
"Calibration" sticker 2.
Verify each quality related measuring instrument2
that will be used during a shift; record the verification and all other
pertinent information on the Measuring
Instrument Verification Log; IF
THE INSTRUMENT AND/OR STANDARD IS DAMAGED, HAS AN EXPIRED CALIBRATION
STICKER, FAILS TO ZERO, OR FAILS VERIFICATION AGAINST A CURRENT
CALIBRATED STANDARD, SUBMIT IT IMMEDIATELY TO THE QC Supervisor OR
CALIBRATIONS CLERK -- DO NOT USE!!
STEP
WHO
WHEN
INFO TO USE
WHAT TO DO
1
Any
team
member
Before
using
any
precision measuring instrument
The calibration
sticker
on
the instrument and the Measuring
Instrument Verification Log
2
QC
Supervisor
or Calibrations Clerk
Instrument(s)
needs to be
repaired or adjusted
Verbal
and/or
visual
information from any team member about the instrument(s) condition
Submit the
instrument(s) to
Metrology; initiate recheck per TOP-RECHK.
3
Any
team
member (typically QC
Supervisor)
Having
any
precision measuring
instrument needing to be placed into the Calibration System
Personal
knowledge
Submit
instrument to the
Calibrations Clerk
for an acceptance review.
4
Calibrations
Clerk
Upon
receiving a precision
instrument for entry or update into the Calibration System
Request
from
submitter
Calibrate
the
instrument per
step 6.
5
Calibrations Clerk
As
instruments become due for
periodic calibration or are submitted for calibration due to
questionable performance
Calibration System
A list shall be pulled at the
beginning of each month reflecting what is due to expire in that given
month and sort by date. This list shall be hung in the QA office for
anyone to see and will also be sorted and sent the respective QC
department, Machine or Forge. This list shall be updated weekly or more
depending on the number of items calibrated throughout the week.
Collect
all
instruments that
are due or that have been turned in by users for calibration; identify
in the Calibration
System
any instruments that have been lost, discarded
or
taken out of service.
6
Calibrations Clerk
As
instruments are submitted
for calibration and as due dates make it necessary
Calibration
System
, ISPEC-CALIB
7 Calibrations Clerk As instruments are submitted for calibration and as due dates makes it necessary. Calibrations Clerk, ISPEC-CALIB For
any gages outside the scope of Trinity's calibration program. An
oustide ISO17025 accredited lab shall be able to calibrate gages per
the ANSI/NCSL Z540 latest revision standard and/or ASME-B-893 standard for software involved calibration, adjustment or repair.
Create a PO for the outside lab ensuring that the vendor is on Trinity's AVL.
Move the gage to "Out for Calibration" in GTS
Properly package the gage and a copy of the PO for delivery to the calibration lab.
8
Calibrations Clerk
For
any
instrument failing
calibration (including vendor calibrations) that cannot be economically
repaired
Results
of
calibration
Tag
the
defective instrument
with a red REJECT label and make the appropriate entry in the
Calibration
System
; return personal or customer instruments to the owner;
properly dispose of Trinity Forge defective instruments; notify the
supervisor(s) of affected area.
9
Calibrations Clerk
For
any
instrument failing
calibration (including vendor calibrations) that can be economically
repaired
Results
of
Calibration
Notify
the Supervisor;.repair
instrument in-house, if possible, or send to an
approved
calibration vendor per step 7 and initiate re-checking per TOP-RECHK;
in the case of Visual Acuity failure, after corrective action is taken,
refer to step 6.
10
Calibrations Clerk
When
an
instrument passes
calibration
Results
of
calibration
Decide
on
the
next calibration
date based on type, condition, environment, use of the instrument and
recommended calibration frequency4;
update the Calibration
System
and tag the instrument with a calibration
sticker
showing
identification, calibration date, next calibration date and the
calibrator’s stamp; apply any safeguard to
protect the instrument from
adjustments that would invalidate the calibration; return the
instrument to user with any recommendations.
11 Calibrations Clerk When
an
instrument passes
calibration from an outside vendor Results
of
calibration; Calibration Cert Review calibration cert for all necessary requirements and that sticker and calibration cert have matching dates.
Move
gage back to "Active" in GTS and calibrate per the provided cert.
(Note: when outside calibration is used, calibration date is the date
provided on the cert from the service provider)
Return gage to service. NO GAGE IS TO BE PUT BACK INTO SERVICE PRIOR TO UPDATING OF GTS
12
Calibrations Clerk
When a
calibration date needs
to be extended
Calibration System
or own judgment
Calibration
due dates may be
extended by no more than one-half of its normal calibration period
under circumstances deemed appropriate; note the extension and
circumstances in the Calibration
System
13
Calibrations Clerk
When
instruments are not in
use for extended periods or are missing
Personal
knowledge
For an
instrument that is not
in use it may be tagged as “IA” (inactive), appropriately stored and
noted in the Calibration
System
; tools that are missing will be entered
into the Calibration
System as such; a “Tool Status” e-mail is
to be sent
to the appropriate supervisor(s).
14
Calibrations Clerk,
or any supervisor
When
an
inactive ("IA")
precision measuring instrument is needed
Personal
knowledge and the
calibration
sticker on instrument
Go to
step 4
unless the
Calibrations Clerk
is not available; in the event that the Calibrations
Clerk is not available, and a tool is needed in an emergency,
a
supervisor may temporarily activate the tool by doing a calibration per
step 6; an email must be sent immediately to Calibrations Clerk; when
the Calibrations Clerk
becomes available, the tool will be submitted
for proper calibration and entry into the Calibration System
.
15
QA Manager
Periodically,
typically monthly
Calibration System;
calibration
stickers
Assign
a Trinity Forge team member to perform a Mini-Audit per TOP-IAUDT
of
the Calibration
System by randomly selecting gages from various shop
locations and tracing their calibration back to the database and
supporting documentation.
2Includes, but is not restricted to any existing calibrated and logged precision measuring device (caliper, CMM, v-block, micrometer, scale, fixture, gage etc.), any precision measuring device not previously entered into the Calibration System , any previously rejected precision measuring device that has been repaired, or any individual required by TOPS (QC Inspectors, Magnaflux Inspectors, etc.) to be entered into the Calibration System.
3Calibration frequency is 12 months for CMM, Surface Tables, Master Blocks, CNC Machines, Thread Gages, and 1-5 years (based on usage) on Standard Blocks. Visual Acuity examinations will be given within the first ninety (90) days of hire date regardless of temporary or full-time employment status. This testing is required for QC Inspectors, QC Leadman, Post Forge Workers that work on the inspection table, or anyone doing a visual inspection of a part as part of final review; this testing will be administered annually; magnaflux Level II every 5 years; calibration frequency on all other instruments is 6 months unless otherwise specified by the QA Manager.4Calibration System includes but is not restricted to Quality Control, the Calibration Log, the Verification Log, and Gage Tracking System (GTS).
Revision history:
Release Date |
Description of Change |
Owner |
Approver |
12/06/2017 |
Updated to comply with revised Quality Manual and new TOPs format. |
Tim Ellis |
Todd Sheppard |
01/18/2023 |
Updated Step 5 to reflect pulling monthly lists for gages that will be coming up for calibration. |
Dustin Jump |
Todd Sheppard |
2/23/2023 |
Updated to better define process for sending out gages for calibration, added steps 7 and 11 |
Dustin Jump |
Todd Sheppard |
|
|
|
|
|
|
|
|