TOP-RECHK: Rechecking Measurements With Out-Of-Calibration Instruments
Rev: 12/06/2017
Scope: This procedure covers the assessment and documentation of the validity of previous inspection and test results when inspection, measuring and test equipment is found to be out of calibration.
Procedure:
*Time
is
critical in rechecking a suspected out-of-tolerance condition. The QA Manager is required to
immediately report any lack of cooperation to
the VP-Quality &
Engineering.
STEP
WHO
WHEN
INFO TO USE
WHAT TO DO
1
Any
team
member
Upon
finding any caliper, gage or other measuring device to be out of
calibration
Own
verification or calibration results
Report
situation to the Calibrations
Clerk*
2
Calibrations Clerk
Upon
receiving report that measuring device is out of calibration
Report
from
team member
Determine
the base point; the most recent time that the faulty device was
calibrated or verified as being in calibration; determine all uses that
device has possibly been used since base point; document possible uses
on Out-of-Calibration
Report; continue to step 3
3
QC Supervisor
Upon
verifying possible use of faulty device
Out-of-Calibration
Report
Direct
the
re-inspection of each application to which faulty device may have been
used, bringing other team members, suppliers or customer in for
information or assistance as needed (see TOP-RCALL);
document the results of each re-inspection on Out-of-Calibration
Report
and produce corrected updates of any records produced from original
readings (if any original results cannot be verified, so note) Return
completed Out-of-Calibration
Report
to the Calibrations Clerk
4
Calibrations Clerk
Receipt
or
production of Out-of-Calibration
Report
Out-of-Calibration
Report
Review
and
confirm full execution of Out-of-Calibration
Report
and confirm in
computer
5
Calibrations Clerk
After
completing rechecks
Out-of-Calibration
Report
Complete Out-of-Calibration
Report and submit to QA
Manager
6
QA Manager
Upon
receiving completed Out-of-Calibration
Report
Out-of-Calibration
Report;
any necessary
quality
records
Review Out-of-Calibration
Report
for clarity and completeness, auditing any
points deemed appropriate; file in Out-of-Calibration
Report
file